Standard formulation unit for Cleanrooms and Related Controlled Environments Part 3: Testing Methods

24-hour service hotline
400-728-7728
KQET-5016 Remote Laser Airborne Particle Counter with Low Flow Rate of 2.83L

The KQET-5016 series is a remote laser airborne particle counter with low flow rate of 2.83L, a testing instrument applied in the fields of cleanliness monitoring and micro-particle counting detection in clean environments.

Based on the light scattering principle, the counter realizes the detection of airborne particle size distribution and the counting of particle quantity through photoelectric sensing technology.

This instrument has been widely used in industries such as cleanroom testing, filter material testing, and mask testing. It features good stability, high measurement accuracy, small size, light weight, and easy installation.

Consultation hotline:181-5178-9348
PRODUCT SPECIFICATIONS


Product Features:

• High-precision sensor, accurate counting, excellent performance, in line with JJF1190-2008 Specification

• Compact size for convenient installation

• Built-in vacuum source with a 3-year long service life, low vibration, low noise, and stable performance to ensure stable counting

• Equipped with a flow sensor for real-time monitoring and automatic adjustment of flow rate, ensuring flow stability and counting accuracy

• 10-year service life imported semiconductor laser diode (calibration required annually)

• RS485 communication, supports Modbus RTU protocol, and customized communication protocols

• Intelligent IoT capabilities allow for remote start/stop, more accurate cleanliness level management, and enhanced monitoring quality

• Expandable analog input hardware interfaces, allowing for external connections to temperature and humidity sensors, airspeed sensors, pressure differential sensors, etc. (optional)

• Real-time monitoring of particle quantities with particle sizes of 0.5μm and 5.0μm; expandable to monitor 6 particle size ranges: 0.3μm, 0.5μm, 1μm, 3μm, 5μm, and 10μm

Technical Specifications:

Model

KQET-5016

KQET-5016M

Particle Size Range

0.3μm~10.0μm

Particle Size Channels

0.5μm, 5.0μm (default)

0.3μm, 0.5μm, 1.0μm, 3.0μm, 5.0μm, 10.0μm

Particle Size Distribution Error

/

≤±30%

Concentration Indication Error

0.5μm≤±20%FS

Maximum Sampling Concentration

35000 particles/L

Repeatability Error

±5%FS

Self-Purification Time

≤5min

Sampling Flow Rate

2.83L±5%

Laser Source

10-year service life imported semiconductor laser diode

Vacuum Source

Built-in vacuum source, 3-year service life

Power Supply

DC24V/1A

Communication Mode

RS485 @ Modbus RTU

Calibration Specification

JJF1190-2008/ISO 21501-4

Enclosure

304 stainless steel

Dimensions (mm)

Width 100 * Depth 158 * Height 68

Operating Environment

Temperature 0~40°C, Humidity 10%~90% (no condensation)

Storage Environment

Temperature -10~50°C, Humidity 0~90% (no condensation)


Contact

Address: 1001, Building 2, No. 209 Zhuyuan Road, Suzhou High tech Zone

Phone: 0512-68561567

Fax: 400-728-7728

Website: www.kqet.com. cn

Email: popularize@kqet.com.cn