Service hotline:400-728-7728
Mobile phone:181-5178-9348
Mailbox:popularize@kqet.com.cn
Address: 1001, Building 2, No. 209 Zhuyuan Road, Suzhou High tech Zone
The 2010 version of GMP management specifications clearly requires that the cleanliness and microorganisms of compressed air must be tested. The KQ-P910 compressed air testing supporting device fully meets the requirements of GMP management specifications and can be adapted to compressed air testing within a pressure range of 30-150psi.
By reducing the pressure of high-pressure gas and performing uniform flow diffusion, the downstream airborne particle counter and bioaerosol sampler can achieve stable sampling.
During the sampling process of this product, no input gas is wasted, and 100% of the input gas is used for sampling. It has wide applicability, safety, and reliability.
• Applicable to any inert gas.
• Simple and fast setup and installation.
• Customizable output flow rate to meet different sampling needs.
• Intuitive user control mode.
• Quiet operation with low noise.
Model |
KQ-P910 |
Adaptable Input Pressure |
0.1-1.0 MPa (30-150 psi) |
Test Particle Size Range |
≥0.3μm |
Compatible Gases |
Nitrogen, argon, helium |
Compatible Instruments |
a. 28.3L/min particle counter; b. 50L/min, 100L/min airborne bacteria samplers |
Dimensions (mm) |
Width 152 * Depth 300 * Height 200 |
Weight |
3.3Kg |
Noise Range |
≤60 db |
Material |
304 stainless steel |
Address: 1001, Building 2, No. 209 Zhuyuan Road, Suzhou High tech Zone
Phone: 0512-68561567
Fax: 400-728-7728
Website: www.kqet.com. cn
Email: popularize@kqet.com.cn